TS 750/TS 150

  • Specifications
  • Documentation
Specifications

Workpiece touch probe

TS 750

TS 150

Probing accuracy

±1 µm with use of the T434 standard stylus
(at a probing speed of 500 mm/min)

±5 µm with use of the T404 standard stylus
(at a probing speed of 1000 mm/min)

Probe repeatability

Repeated probing from
one direction

2 s 0.25 µm

2 s 1 µm

Probing force

Axial:  1.5 N

Radial:  0.2 N

Axial:  7 N

Radial:  1 N

Deflection of probe contact

5 mm in all directions (with stylus length L = 40 mm)

Deflection forces

Axial:  8 N

Radial:  1 N

Probing speed

3 m/min

Protection EN 60529

IP68

Operating temperature

10 °C to 40 °C

Storage temperature

–20 °C to 70 °C

Mass without base

 0.1 kg

 0.1 kg

Fastening*

  • Base with M22x1 thread and axial cable outlet
  • Base with three M3 screws and axial cable outlet
  • Base with four M3 screws and axial cable outlet
  • Optional: touch-probe extension with M16x1 thread

Electrical connection*

Two-pole sliding contact on the mounting base

Signal transmission

Cable: ≤ 25 m

Supply voltage1)

DC 15 V to 30 V at 85 mA (without load)

Output signals1)

Trigger signals S and S (square-wave signal and its inverted signal)

Floating trigger output

HTL signal levels1)

UH 20 V at –IH 20 mA

UL 2.8 V at –IL 20 mA

at rated voltage of DC 24 V

* Please select when ordering

1) Via UTI 150